Design and preliminary test of liquid phase pVT measurement system using constant-volume cell

Bao-l AN,Yuan-yuan DUAN,Long-shan TAN,Zhen YANG
DOI: https://doi.org/10.13738/j.issn.1671-8097.2014.04.002
2014-01-01
Abstract:Liquid phase pVT behavior is an important thermodynamic property and can be measured precisely with a constant-volume cell. Based on the fundamental theory of the constant-volume method, a new apparatus for liquid phase pVT measurement was designed, including a constant-volume cell,measurement systems of temperature,pressure,mass and a vacuum system. The volume of the constant-volume cell was calibrated with high purity R134a at a temperature of 294.899 K and a pressure of 21.847 MPa.Then the effect of temperature and pressure on the cell volume was analyzed.The pVT behavior of R134a in liquid phase was measured at a temperature range from 270 K to 295 K and pressure from 1.7 MPa to 21.8 MPa.The experimental results agree well with the reference equation of state,which demonstrates the reliability and accuracy of the apparatus.
What problem does this paper attempt to address?