Abstract:A precise pressure-volume-temperature(PVT) measurement system for refrigerants was developed,which was controlled automatically by a computer,the software used to measure and control the system was programmed.The analysis results indicated that temperature uncertainty of PVT measurement system was less than ±1.3 kPa.The measurement chamber volume was calibrated by gaseous helium.The PVT properties of HCFC22 at 293.15~323.15 K were measured to verify the accuracy of the system.The results showed that the average deviation of density between the experimental data and the data provided by the National Institute of Standards and Technology(NIST) was below ±0.064%,and that the maximum deviation was below ±0.356%.This system has advantages of high automation,high stability,and credibility.