Determination of pseudoephedrine in human serum by HPLC method

LIU Gang,WANG Hui,ZHANG Xian-zhou,SONG Jin-chun
2006-01-01
Abstract:OBJECTIVE To establish a RP-HPLC method for determining the concentration of pseudoephedrine in human serum.METHODS Zorbax SB-18 column((4.6) mm×250 mm,5 μm) was used,(0.2%) phosphoric acid solution-acetonitrile(95∶5) were served as mobile phase.The detection wavelength was 194 nm.RESULTS Pseudoephedrine concentration presented a good linear range of 50~(1 000) μg·L~(-1)(r=(0.999 3)),the average recovery was(79.20%).The relative standard deviations of within-day and between-day were less than 5%.The minimal detectable concentration in plasma was 25 μg·L~(-1).CONCLUSION This method is rapid,simple and accurate for the determination of pseudoephedrine levers in human plasma.It was suitable for the pharmacokinetics and bioavailability study of pseudoephedrine.