Detrmination of Pseudoephedrine in Human Serum by HPCE

Liu Gang,Wang Hui,Zhang Xianzhou
DOI: https://doi.org/10.3969/j.issn.1008-049X.2006.07.016
2006-01-01
Abstract:Objective: To establish a HPCE method for determining the content of pseudoephedrine in human serum. Method: HPCE was used with the capillary column(45 cm×50μm). The mobile phase was 0.05 mol·L-1 boronic acid buffer. The applied voltage was 20 kV at 25℃. The eletroinjection was 3. 15 Pa for 15 s and the detection wavelength was at 210 nm. Result: Linear range was 50-500 ng·ml-1(r =0.999 6) ,the average recovery was 79.20% . The relative standard deviations of within-day and between-day were less than 5%. The minimal detectable concentration in plasma was 25 ng·ml-1. Conclusion: This method is rapid, simple and accurate for the determination of Pseudoephedrine levers in human plasma. It was suitable for the pharmacokinetics and bioavailability study of Pseudoephedrine.