Determination of pseudoephedrine in the human plasma by liquid chromatography-tandem mass spectrometry

杨汉煜,陈笑艳,徐海燕,侯悦,钟大放
DOI: https://doi.org/10.3969/j.issn.1006-2858.2001.02.013
2001-01-01
Abstract:A specific,sensitive and rapid LC/MS/MS method for determinationof pseudoephedrine in the human plasma was developed. Pseudoephedrine and the internal standard ropivacaine were extracted from plasma using liquid-liquid extraction,then separated on a Zorbax C18 column. The mobile phase consisted of methanol-water-formic acid (60∶40∶2,V/V),and was set at a flow rate of 0.2 mL/min. A Finnigan TSQ tandem mass spectrometer equipped with electrospray ionization source was used as the detector and was operated in the positive ion mode. Selected reaction monitoring (SRM) using the precursor→product ion combinations of m/z 166→147 and m/z 275→125 was used to quantify the pseudoephedrine and the internal standard,respectively. The linear calibration curves were obtained in the concentration range of 5.0~250.0 ng/mL. The limit of quantification was 5.0 ng/mL. Within-day and between-day precision were within 6.7% and 8.3%,respectively. The method was successfully applied to determining the concentrations of pseudoephedrine in human plasma after an oral administration of 60 mg pseudoephedrine hydrochloride to 20 healthy volunteers.
What problem does this paper attempt to address?