Measurement and analysis of setup error in intensity modulated radiation therapy for head-neck tumor

SHEN Jun-Shu,WANG Jian-Qi,ZHANG Yi-Qin,ZHAI Zhen-Yu,JING Xi-Jing
2006-01-01
Abstract:AIM: To measure the setup error in the intensity modulated radiation therapy (IMRT) for head-neck neoplasms, and to analyze the margin between the clinical target volume (CTV) and planning target volume (PTV). METHODS: Seventy -six patients with head-neck neoplasms were choosed at random. The portal films were screened using the electronic portal imaging device (EPID), and the error was analyzed through comparing the portal film with the digitally reconstruced radiographs (DRR) in the treatment planning system (TPS). RESULTS: The setup errors were (-0.62±1.46) mm,(-0.41±1.54) mm,(-0.31±1.67) mm and the margins were 2.27 mm, 1.87 mm, 1.98 mm in lateral, cranial-caudal and ventral-dorsal directions respectively. CONCLUSION: For the patients with head-neck neoplasms undergoing the IMRT, the margins between the CTV and the PTV should be 2.5 mm in the lateral direction, and 2 mm in cranial-caudal and ventral-dorsal directions. Moreover, with the treatment time goes by, the error varied irregularly.
What problem does this paper attempt to address?