Effects of the Setup Error on the Dose Distribution for Nasopharyngeal Carcinoma with Intensity Modulated Radiation Therapy

Qingguo Xie
2010-01-01
Abstract:Objective To evaluate the effect of the setup error on the physics dose distribution of targets and peripheral organs at risk(OAR) undergoing the intensity modulated radiation therapy for nasopharyngeal carcinoma(NPC).Methods Six NPC patients were included into our study.The On-Borad-Imager of Varian Trilogy linear accelerator was used to collect the images of Cone Beam Computerized Tomography(CBCT) in real time,which were matched with the positioning CT images before radiation treatment through target center.The Varian software was used to calculate the error in x(left-right),y(cranial-caudal) and z(ventral-dorsal) direction.Then the treatment planning system of Eclipse8.6Version was used to study the effects of the setup error on the dose distribution undergoing the radiation therapy for NPC.Results The systematic±random error in x,y,and z direction was(-0.52±1.20)mm,(-0.22±2.54)mm and(-0.53±1.21)mm,respectively.The extent of dose variation for GTVnd,GTVnx,left lens,right lens,spinal cord,and brainstem was 0.32%~12.34%,-0.25%~0.32%,-12.45%~0.13%,-17.87%~-0.08%,-0.18%~8.20% and-5.40%~6.55%.Conclusion The setup errors were inevitable in radiation therapy.However,the OBI-CBCT system could revise the setup error in real time,which could reduce the systematic and random error and improve the accuracy of setup and dose of targets and OAR.As a result,it could provide better quality assurance in radiation therapy.