Setup error analysis for cervical cancer in image guided intensitymodulated radiotherapy

Ransheng LIU,Bo JIANG,Xiangli SHI,Wei WANG,Zhiyong YUAN
DOI: https://doi.org/10.3969/j.issn.1009-6469.2017.02.030
2017-01-01
Abstract:Objective To investigate the importance of cone-beam computed tomography (CBCT) scan in setup of intensity-modulated radiation therapy (IMRT) for cervical cancer.Methods 20 patients with cervical cancer were enrolled into the study and were immbolized with abdomen board and thermoplastic.All the patients were scanned by kilo voltage cone-beam computed tomography before radiotherapy every week.The setup errors isocenter positions on the axes of X,Y and Z were record and the data and correct placement error were analyzed online.Results 20 patients (120 cases) were verified about the position by CBCT.Before correction,the setup error of 3 directions (X,Y,Z axis) is (0.295±0.074),(0.373±0.090) and (0.302±0.078),respectively;After the correction,the setup error is (0.014±0.012),(0.034±0.022),(0.020±0.017) cm,respectively.All patients were successfully treated and the setup error decreased gradually.One year local tumor control rate was 100% and the adverse reaction decreased significantly.Conclusion CBCT scan is useful to analyze the causes of errors,taking effective measures to reduce the setup errors can make the does distribution of target and normal tissues around the organs more accurate,which provide quality assurance for clinical radiotherapy.