A Clinical Evaluation of Setup Errors in IMRT Using Cone Beam CT for Head and Neck Cancer

廖希一,林少俊,吴君心,潘才佳,柏朋刚,潘建基
2013-01-01
Abstract:[Purpose]To investigate the linear and rotational setup errors in intensity modulated radiation therapy (IMRT) using kV cone beam CT (CBCT) on IGRT for head and neck cancer. [Methods] Twelve patients with head and neck carcinoma were treated by Elekta Synergy IGRT with a total of 217 times.X-ray volumetric images(XVI) acquired pre-,intra-(after correcting the setup error),and post-radiation,were compared with planned CT images to get informations of linear and rotational setup errors of X-axis,Y-axis,Z-axis.[Results] A total of 217 CBCT scans were applied for 12 patients pre-radiation.The systemic±random errors on linear X,Y,Z directions were (-0.04±2.63)mm,(0.07±1.69)mm,(-1.15±1.33)mm,on rotational X,Y,Z directions were (0.10°±1.14°)、(0.16°±1.41°)、(-0.06°±1.22°)respectively. After correction of all the setup errors,both the systematic errors and random errors were obviously less than those before correction.The difference of setup errors between intra-radiation (after correcting the setup error) and post-radiation was not significant. [Conclusion] Real-time correction of setup error in patients with head and neck cancer by CBCT can reduce the setup error and decrease planning target volume (PTV) extending. And the setup error in fractionated radiation for head and neck cancer is small.