EPID for Set-up Errors Analysis and Quality Assurance of Radiotherapy in Head and Neck Cancer

Chang-hua YU,Ji-hua HAN,Zhen-ya ZHU,Xiao-ye ZHANG,Wei-guo ZHU,Tao LI
DOI: https://doi.org/10.3969/j.issn.2095-2457.2015.17.224
2009-01-01
Abstract:Objective: This study was to investigate the set-up error using electronic portal imaging device (EPID) for intensity modulated radiation ( IMRT), assess of the CTV-to-PTV margins, analyze the factors on causing the error periodically, and to improve the quality control level. Methods: From October 2007 to January 2008, twelve, eleven, seventeen, and sixteen patients with head and neck carcinoma were enrolled in this study. The patients were immobilized with the head/neck shoulder immobilization system on margins. The set-up errors were analyzed by comparing the films of digitally reconstructed radiographs (DRR) with EPID. Electronic portal imaging was performed for twice each week before each treatment at the position AP (gantry angle 0°) and LT (gantry angle 90°), respectively. CTV-to-PTV margins were calculated by the equation 2.0Σ+0.7σ (Σ is the systematic error, and σ is the random error) in the end of each month, the factors which might be the causes were analyzed and some quality control methods were proposed. Results: The theoretical CTV-to-PTV margins calculating by the equation was declined gradually during the previous three months and stable in the fourth month. Conclusions: This study supplied theoretical evidences and validated CTV-to-PTV margins for our department of radiotherapy by analyzing the set-up errors. This strategy may diminish the unnecessary irradiation fields and enhance the treatment outcome through the technique improvement.