Setup Verification Using Electronic Portal Imaging for Nasopharyngeal Carcinoma Intensity Modulated Radiation Therapy

Pan Caizhu,Pan Jianji,Chen Chuanben,Zhang Xiuchun,Li Jinlian
DOI: https://doi.org/10.3969/j.issn.1672-4194.2007.01.014
2007-01-01
Abstract:Objective To investigate the setup verification using electronic portal imaging for nasopharyngeal carcinoma intensity modulated radiation(IMRT).Methods Thirty six patients were enrolled in study.Electronic portal images(EPI) were acquired once every week in the radiation course and a total 372 anterior or lateral isocentric EPI were compared to digitally reconstructed radiograph(DRR) using electronic portal imaging device(EPID).Results About 90.6% of all translational setup errors were ≤3 mm,and 99.2%≤5 mm.The overall systematic error was-0.5 mm(±1.2 s) in X-axis(right-left direction),-0.5 mm(±1.4 s) in Y-axis(superior-inferior direction) and-0.9 mm(±1.5 s) in Z-axis(anterior-posterior direction).The corresponding s of the random errors were ±0.3,±0.2 and ±0.3 mm.The estimated margins required for CTV-PTV were calculated according to the van Herk formula was 3.2,3.6 and 4.0 mm,respectively.Conclusion There exists some extent of setup error in nasopharyngeal carcinoma IMRT.On-line correction of setup error using EPID in order to reduce systematic error and random error and then improve setup precision.