Evaluation of Set-up Errors for 493 Patients by Simulator

Hao Huang
2011-01-01
Abstract:Objective: To evaluate set-up error by Varian Acuity simulator.Methods: From October 2009 to December 2010,493 patients were included in this study,including 167 receiving head and neck irradiation,210 thoracic irradiation and 116 abdominal irradiation.All patients were immobilized with fixation devices.DRRs of each patient were compared with portal images from simulator and set-up errors were recorded and analyzed.The shift margin from CTV or ITV to PTV was calculated.Results: In head and neck region,the set-up errors in Right-Left,Cranial-Caudal and Anterior-Posterior directions were(0.13±0.17)cm、(0.21±0.27) cm and(0.21±0.23)cm.In thoracic region,the set-up errors in R-L,C-C and A-P directions were(0.24±0.27)cm、(0.24±0.32) cm and(0.27±0.29)cm.In abdominal region,the set-up errors in R-L,C-C and A-P directions were(0.22±0.29)cm、(0.25±0.29)cm and(0.19±0.29)cm.And the margins from CTV or ITV to PTV were given.Conclusions: Set-up errors and margins from CTV or ITV to PTV can be obtained by simulator.