Kilo-volt Cone-beam CT Evaluation of Set-up Errors in Nasopharyngeal Carcinoma Treated with Intensity Modulation Radiotherapy and Its Effects on Dose Distribution

徐升,张军宁,周菊英,符天晓,徐久宏,陈龙,陈浩
DOI: https://doi.org/10.3969/j.issn.1001-5930.2014.09.048
2014-01-01
Abstract:Objective Objective To evaluate the set-up errors of intensity-modulated radiotherapy ( IMRT) in nasopha-ryngeal carcinoma ( NPC) and analyze its effects on dose distribution of targets and peripheral organs at risk ( OAR) using kilo-volt cone-beam CT ( KV-CBCT) .Methods 15 NPC patients who received IMRT were selected .They were set up by 3 points on thermoplastic film ,and scanned by KV-CBCT once a week .Sectional images were acquired after reconstruction and compared with the planning CT images ,the variations of three dimensional and horizontal rotation could be achieved .Transplanting the offset cen-tral point of three dimensional and horizontal rotation to the primary positioning images ,the other parameters were invariant .The dose distributions were simulated on the primary treatment planning CT images and the dose distributions betweem them were compared.Results The mean values of set-up errors on anteroposterior,craniocaudal,lateral position and rotation was -1.833 mm,-0.32 mm,0.96 mm,-0.8333 degree,respectively.The dose variations for GTVnxD95,GTVndD95,CTV2D95,PTV2D95,and PTVnd D95 were -1.28%~-0.32%,-3.31%~-0.63%,-2.33%~-1.30%,-2.51%~-0.45%,and -2.53%~-0.48%,respectively.Dmax of brainstem,spinal cord,and the optic chiasm were -3.14%~12.58%,-22.91%~21.21%,-7.25%~14.77%,respectively.The D50 and Dmean of parotid were -13.87%~18.21%and -6.84%~9.21%.Conclusion Due to the presences of set-up errors,dose distribution of nasopharyngeal carcinoma ( NPC) undergoing IMRT is affected ,and dose distribution of OAR increased significantly .KV-CBCT can reduce the treatment errors and improve the accuracy of IMRT .
What problem does this paper attempt to address?