Improvement of a Stirling type two-stage pulse tube cryocooler

Yan Pengda,Chen Guobang,Dong Jingjing,Gao Weili
DOI: https://doi.org/10.3969/j.issn.1000-6516.2008.05.002
IF: 2.134
2008-01-01
Cryogenics
Abstract:Based on the preliminary test of two-stage pulse tube cryocooler driven by linear compressor,the coil of the compressor was modified,to input enough electric power.A slot type water cooled heat exchanger was adopted and built in the warm end flange,so that the heat exchange efficiency may be improved and the overheating problem of the warm end was solved.The performance of cryocooler was improved obviously.A lowest refrigeration temperature of 14.2 K was achieved and two cooling capacity of 2.5 W at 97.8 K and 1 W at 34.9 K were demonstrated at the first stage and second stage respectively.