Benefits of a special cooling system to improve precision and accuracy in non-vaporizing on-column injection procedures

M. Galli,S. Trestianu
DOI: https://doi.org/10.1016/s0021-9673(00)80293-9
IF: 4.601
1981-01-01
Journal of Chromatography A
Abstract:The use of a special cooling system to cool the first part of the capillary column during rapid, non-vaporizing on-column injection eliminates or at least drastically reduces the consequences of the sample back-returning process. This process produces sample loss, discrimination and injector contamination, not only at oven temperatures around or above the boiling point of the solvent, but also at lower temperatures. The results obtained show that the cooling system is successful in keeping the sample inside the capillary within a reasonably wide range of operating conditions, thereby improving the precision and accuracy of the chromatographic data.
chemistry, analytical,biochemical research methods
What problem does this paper attempt to address?