The Design of Resistive-Heating System for Capillary Column

ZHANG Xuanlang,ZOU Yingchang,ZHANG Xi,CHEN Chenxing,AN Chao,WANG Ping,CHEN Xing
DOI: https://doi.org/10.3969/j.issn.1004-1699.2013.07.002
2013-01-01
Abstract:This article introduces a design of resistive-heating system for capillary column used in fast gas chromatography(GC).Direct Current was applied to the capillary column for heating.3-wire platinum resistor and constant current source were used for the temperature measurement.The heating power was controlled by a PWM actuator using a PID algorithm.Eventually the heating system can heat the capillary column from room temperature to 200 ℃ at a rate from 1 ℃/s to 10 ℃/s.The temperature precision is 0.2 ℃.And the system is portable and low power consumption,which can meet the requirement for fast GC.
What problem does this paper attempt to address?