Fundamental and practical considerations in the design of on-line column preconcentration for flow-injection atomic spectrometric systems

Zhaolun Fang,Shukun Xu,Suchun Zhang
DOI: https://doi.org/10.1016/S0003-2670(00)83756-3
IF: 6.911
1987-01-01
Analytica Chimica Acta
Abstract:Fundamental consideration and practical points in the development of on-line preconcentration flow-injection with atomic spectrometric detection are discussed, with emphasis on overall efficiency and accuracy. The terms concentration efficiency (CE) expressed as the enrichment factor achieved per minute by the system, and retention efficiency (%E), together with recovery studies on real samples, are recommended as criteria for the evaluation of the efficiency and reliability of such systems. Time-based sampling and double-column systems without intermediate column washing are recommended for improving efficiency. Large columns (3 mm i.d., 45 mm long) with high sample-loading rates of 9.5 ml min−1 are proposed for achieving high efficiency and accuracy with atomic absorption detection; more careful optimization of column dimensions is needed for detection with inductively-coupled plasma/atomic emission spectrometry. A procedure for the on-line preconcentration of cobalt in water samples was developed under the guidelines presented. An enrichment factor of 48 was achieved at a sampling frequency of 60 h−1 with good recoveries for all the water types studied. The precision was 1.7% r.s.d. at the 40 μg1−1 level, and the detection limit (3σ) was 0.2 μg1−1.