Modeling and Analysis of Flow Through the Orifice of Still in Dilution Refrigerator

De-Ming Wang,Wei-Pei Cha,Wei-Dong Ding,Dong-Ming Xu,Wan-Long Han,Jian-Xin Lu,Qin Zhan
DOI: https://doi.org/10.1007/s10909-024-03157-3
2024-05-30
Journal of Low Temperature Physics
Abstract:The cooling capacity of a typical dilution refrigerator is mainly determined by the circulating flow of the He-3 in the system, which is usually realized by the external circulation pump. In order to increase the concentration of circulating He-3, the still often take the orifice structure to limit the influence of the superfluid helium film, and this structure will also have a certain impact on the flow rate of He-3. This paper summarizes the flow rate models of each part of the orifice structure, and establishes a model related to the influence of pumping speed and the diameter of orifice on the flow rate of circulating He-3. The relationship between pumping speed and flow rate is developed and the relation is divided into three phases for detailed analysis. It is found that the flow rate and the concentration of circulating He-3 increase first and then decrease with the increasing diameter of orifice at a certain pumping speed, and the possible reasons for this phenomenon are analyzed. The results indicate that there is an optimal orifice diameter at a determined pumping speed.
physics, condensed matter, applied
What problem does this paper attempt to address?