Study on the Calibration of Partial Discharge Calibrators of Charge from 0.1pc to 10Nc

Na Liu,Shengyou Gao,Bo Li,Dongxu
DOI: https://doi.org/10.1109/cpem61406.2024.10646169
2024-01-01
Abstract:This paper describes NIM’s measurement methods for Partial Discharge(PD) calibrators of 10nC and below 1pC in 2023 for participation of supplementary comparison EURAMET.EM-s36. Because the comparison range is from 0.1pC to 10nC, which is very wide. The impulse voltage signals are from a few millivolts to tens of volts, therefore different measurement methods are applied by NIM for low signal levels and strong signal levels in order to obtain low measurement uncertainty. It is difficult for the numerical integration method recommended in IEC60270-2015. Therefore, in this work the measuring resistor is not a single low-inductance resistor. But a set of low-inductance resistors ranging from 5Ω to 2kΩ are developed. For charge below 1pC, a voltage preamplifier is applied to amplify and improve the signals.