A Study of Calibration System of DC Digital Nanovoltmeter

HU Yi-fei,ZHAO Wei,ZHU Zhu,KANG Yan,HUANG Jing,LI Xiao-yan
DOI: https://doi.org/10.3321/j.issn:1000-1158.2007.01.002
2007-01-01
Abstract:Based on the low thermal EMF divider,a low DC voltage calibration system is developed for DC digital nanovoltmeter calibration.To reduce thermal EMF in measurement circuit,the techniques of connection among cables by special press and temperature-controlled box are studied.The Hamon junctions are fabricated in order to reduce resistance of the leads.The symmetric observation method and linear interpolation method are put forward to reduce the influence of thermal EMF by one order of magnitude.The low DC voltage calibration system shows measuring range from 1 μV to 100 mV,and measurement uncertainty of 2.2% at 1 μV value.