Progress on a Simultaneous Photon Counting and Energy Integrating Detector for X-ray Imaging Applications

X. Ran,Z. Deng,H. Gao,X. Wang,Y. Song
DOI: https://doi.org/10.1109/nss/mic/rtsd57108.2024.10655152
2024-01-01
Abstract:The photon counting X-ray imaging system offers high spatial resolution, low noise, and reduced radiation exposure, but suffers degradation at high counting rates. We have developed a dual mode imaging readout chip suitable for cadmium telluride or cadmium zinc telluride crystals, capable of simultaneous integration and photon counting. Utilizing this chip, we designed an X-ray imaging camera system comprising four detector modules, each with a $32 \times 32$ array of readout channels covering a 16 mm by 64 mm imaging area in total. Electronic testing revealed the noise level of each counting channel in these modules to be approximately 860 electrons, with the integration channel’s leakage current measured at about 1 nA at $20^{\circ} \mathrm{C}$. After correcting the non-uniformity for baseline and gain in the counting channels and the non-uniformity between the qpump and ADC conversion in the integration channels, imaging experiments were performed. Analysis of the concurrently gathered photon counting and integration data demonstrated a correlation between the dual-mode outputs. The system’s dual-mode imaging capabilities were validated by successfully reconstructing a basic water phantom computed tomography image using the filtered backprojection algorithm.
What problem does this paper attempt to address?