Development Of A Multi-Energy Photon-Counting Readout Asic For Czt Detectors

Zhi Deng,Xuezhou Zhu,Yu Chen,Yinong Liu,Yuxiang Xing,Chuqing Feng,Guangqi Wang,Hui Liu,Sen Wu
DOI: https://doi.org/10.1109/NSSMIC.2015.7582265
2015-01-01
Abstract:The paper presents the development of a multienergy photon-counting readout ASIC for CZT detectors for spectral X-ray imaging applications. It integrated 32 identical channels of signal processing circuits, including a charge sensitive preamplifier, a high-order CR-(RC) 6 shaper with 100 ns peaking time and five discriminators with fine threshold adjustment and corresponding window logics and 16-bit counters. The ASIC was optimized for linear CZT detectors with millimeter to submillimeter size and was fabricated in a 0.18 mu m CMOS process. The first prototype was implemented and verified in two separated parts: the analog front-end and the discriminators with counters. Preliminary test results showed the ASICs worked well. The overall gain of the analog front-end was measured to be 170260 mV/fC and the ENC was measured to be < 400 electrons with similar to 3 pF input capacitance. The discriminator and counting function were also verified up to similar to 1 MHz, including the fine adjustment of the thresholds and the data readout. Detailed circuit design and test results will be depicted in the paper.
What problem does this paper attempt to address?