Design and Optimization of Direct-Conversion Photon-Counting Detector for Dual-Energy Ct Imaging

Yannan Jin,Geng Fu,Hewei Gao,Peter M. Edic
DOI: https://doi.org/10.1109/nssmic.2016.8069518
2016-01-01
Abstract:A direct-conversion photon-counting detector has great potential in CT imaging and has been intensively studied in recently years. However, the optimization of the detector configuration based on clinically-relevant metrics is still lacking. In this study, we developed an optimization scheme for CZT-based photon-counting detector design. The spectral response functions characterizing different detector designs were generated using a Monte Carlo-based approach and evaluated with a dual-energy analysis tool. The results demonstrate the degradation in dose efficiency caused by charge sharing when the detector pixel size is below ∼500 μm. The results also show that thinner CZT sensor material has degraded performance due to the lower absorption efficiency. Based on this analysis, ∼500 μm pixel size and 2.5 mm thickness seem to be a good balance between dose efficiency and spatial resolution. The results also show that electronic noise has limited impact on dose efficiency and is not a major risk to the detector development.