Equivalent Circuit Model Of Four-Terminal-Pair Capacitor Based On High-Frequency Traceable Measurement

Dx Dai,Ys Ruan,Q Wang
2004-01-01
Abstract:In order to calibrate four-terminal-pair LCR impedance measurement instrument or impedance/network analyzer in the working frequency range, a set of traceable four-terminal-pair standard capacitors are required. Firstly, the paper briefly describes the calibration theory and the implement procedure of four-terminal-pair air dielectric capacitor and provide the frequency characteristic of measured four-terminal-pair air dielectric standard capacitor, 1000pF, 100pF, 10pF, 1pF from 1 kHz to 13 MHz. Secondly, measurement methods for the equivalent circuit component parameters of four-terminal-pair air dielectric capacitor are described in detail. Finally, some main factors that lead to measurement error of four-terminal-pair capacitor are analyzed. By new and simple measuring methods, the equivalent circuit components of four-terminal-pair capacitor are firstly established. According to the established equivalent circuit model of four-terminal-pair capacitor, which combines to four-terminal-pair admittance definition and the characteristic of four-terminal-pair LCR impedance measurement instrument, it is very helpful to analyze some factors that affect frequency characteristic and measurement error of four-terminal-pair capacitor.