Clinical application of comparison of hematology analyzer by using fresh whole blood transfer method

Yang XU,Yahui ZHENG,Lei ZHANG,Yanping ZHANG,Wen ZHANG
DOI: https://doi.org/10.13201/j.issn.1004-2806-b.2017.10.007
2017-01-01
Abstract:Objective:To study the clinical application of comparison of hematology analyzer by using fresh whole blood transfer meyhod.Method:Fresh whole blood samples from patients were measured by the No.1 hematology analyzer of the Sysmex HST-302 hematology analyzer assembly line which was selected as the reference analyzer.The results of WBC,RBC,HB,PLT and HCT were used to compare and calibrate the No.2 analyzer which belonged to the same assembly line.Transfer the fresh whole blood between two hematology analyzers,the deviation could be found in time.Result:The consistency was confirmed after comparison between two hematology analyzers.And the fresh whole blood transfer method could apply to internal quality control.Conclusion:Through calibration of correct deviation of the two analyzers by using fresh whole blood,the consistencies for different instruments are improved.It is an economical,practical and effective method for clinical application.