Performance Evaluation of Mindray BC 5800 Automatic Hematology Analyzer

MA Yan,CHENG Zhao-min,KE Pei-feng,ZHENG Song-bai,XIONG Xiao-quan,WU Xin-zhong
DOI: https://doi.org/10.3969/j.issn.1674-1633.2012.08.041
2012-01-01
Abstract:Objective To evaluate the performance of Mindray BC 5800 automated hematology analyzer.Methods According to the evaluation standard esdablished by International Committee for Standardization in Haematology(ICSH),to measure the inaccuracy,carryover and linear range of Mindray BC 5800 automatic hematology analyzer,and compare the classified counting results with classified counting and type checking for manual microscope of Sysmex XE 2100.Results The accuracy,liner and carryover of Mindray BC 5800 automatic hematology analyzer are all within allowed range,and its correctness and comparability are good.The differential count relativity r2 of white blood cell and Sysmex XE 2100 is larger than 0.8.Comparing with artificial microscope method,relativity r2of the rest exceeds 0.9 except that monocytes relativity r2is smaller than 0.8.Conclusion Mindray BC 5800 blood cells analyzer is an ideal blood cells analyzer,and can meet clinical demand.
What problem does this paper attempt to address?