Performance evaluation of WBC and RBC counting with the Beckman-Coulter Z2 cell counter

PENG Liming,WANG Weixin,BAI Lixia
IF: 1.865
2007-01-01
Laboratory Medicine
Abstract:Objective To evaluate the performance of the Beckman-Coulter Z2 cell counter for white blood cell(WBC) and red blood cell(RBC) counting. Methods The performance of the Z2 cell counter such as accuracy,precision,linearity,interference-resistance and coincidence counting was evaluated according to guidelines published by the International Committee for Standardization in Hematology(ICSH) and compared with that of Sysmex F-820 and SE-9500. Results The results demonstrated excellent linearity for WBC and RBC(r>0.999).The Z2 showed excellent precision,with within-run,between-run and overall coefficient of variance(CV) less than 2%.Carryover was generally acceptable(<1%).WBC and RBC were stable after blood storage for 48 h at 4 ℃ and room temperature(RT).There was minimal interference for WBC,RBC at high concentrations of bilirubin or triglyceride(P>0.05).The bias of coincidence correction in WBC and RBC counting with the Z2 was less than 1%.Correlation between the Z2 and Sysmex F-820 or SE-9500 was excellent for WBC and RBC counting(r>0.996). Conclusions The Beckman-Coulter Z2 cell counter has good features of precision,accuracy,linearity,interference and small deviations of coincidence counting and good comparison with Sysmex F-820 and SE-9500,and it could be used to other hematology analyzer for WBC and RBC counting.