Establishment and evaluation on the microscopy rule for nucleated red blood cell by SYSMEX XN-9000 automatic hematology analyzer

Meixiu GU,Zhimin MENG,Wenjia TANG,Lei LUO,Chong WANG,Wei GUO,Baishen PAN,Beili WANG
DOI: https://doi.org/10.3969/j.issn.1673-8640.2017.06.019
IF: 1.865
2017-01-01
Laboratory Medicine
Abstract:Objective To investigate the microscopy rule for nucleated red blood cell(NRBC) by SYSMEX XN-9000 automatic hematology analyzer,and to evaluate the correction formula for white blood cell(WBC) count and NRBC.Methods Samples were determined by SYSMEX XN-9000. The results must be >0 NRBC/100 WBC. Totally,716 samples were collected. Every sample was observed by microscopy after Wright-Giemsa staining. The accuracy of SYSMEX XN-9000was evaluated,and the results were compared with those of microscopy(gold standard). The correction formula was investigated.ResultsThe correlation coefficient(r2)between SYSMEX XN-9000 and microscopy was 0.981. Bland-Altman plot showed a good consistency between the 2 methods. The 95% agreement limit of difference between the 2 methods was(-0.59±2.73) NRBC/100 WBC,and 23 samples (3.21%) was beyond the 95% agreement limit of difference. Receiver operating characteristic(ROC)curves were used to identify the optimal smear threshold value. The sensitivity was 87.3%,and the specificity was 76.5% at≥1.1 NRBC/100 WBC,while the area under ROC curve was 0.88. However,the false negative rate was 23.5%,which was against the criterion of ISO 15189(≤5%). At ≥0.3 NRBC/100 WBC,the false positive rate was 49.6%,and the false negative rate was 2.3%. If WBC count error between SYSMEX XN-9000 and microscopy was >5% [1/3 allowable total error(TEa)],the results of SYSMEX XN-9000 should be recalculated according to those of microscopy. The correction formula was WBC=(100×WBCSYSMEX XN-9000+NRBCSYSMEX XN-9000×WBCSYSMEX XN-9000)/(100+NRBCmicroscopy).Conclusions SYSMEX XN-9000 could determine NRBC in peripheral blood rapidly, accurately and effectively. WBC count should be recalculated,when there is difference between SYSMEX XN-9000 and microscopy.
What problem does this paper attempt to address?