The Microscopic examination of abnormal blood pictures displayed by Sysmex XS-1000i

MO Heguo,SUI Hong,XU Ning,CHEN Guanghui,GUO Zhouqing
DOI: https://doi.org/10.3969/j.issn.1674-4721.2011.21.006
2011-01-01
Abstract:Objective:To show the importance of manual morphological examination review when there are abnormal pathological conditions.The Sysmex XS-1000i hematology analyzer can only be used as a screening tool in cell morphology examination.Methods:The oil microscope was used in the observation of cell morphology and manual WBC differential analysis in 1 080 cases which were shown abnormal in the Sysmex XS-1000i hematology analyzer,then the main disease-related form of abnormal cytology group were statistically analyzed.Results:782 cases of abnormal cell morphology can be detected(true positive) in the recheck of 1 080 cases.The detection rate(true positive rate) was 72.40%.Conclusion:The manual morphological examination can not be replaced by Sysmex XS-1000i hematology analyzer especially when there are abnormal blood pictures.Both methods should be used in order to alleviate mistakes and missed diagnosis.
What problem does this paper attempt to address?