Performance evaluation the new automated Atellica Hema 580 hematology analyzer

E. Schapkaitz,A. Baiden,S. Raburabu
DOI: https://doi.org/10.1111/ijlh.14170
2023-09-22
International Journal of Laboratory Hematology
Abstract:Introduction The Atellica Hema (Siemens Healthineers, Tarrytown, NY, USA) is a new generation multi‐parameter analyzer for full blood count, 6‐part differential and reticulocyte testing by impedance variation and fluorescence flow cytometry. In this study, we verified the whole blood and limited body fluid modes of the Atellica Hema 580. Methods We evaluated precision, linearity, carry‐over, throughput and performed a method comparison to assess the performance of the Atellica Hema 580. For comparison of the Atellica Hema 580 with the Sysmex XN‐1000 (Sysmex, Kobe, Japan), 140 samples from adult and pediatric patients including both normal and abnormal hematology profiles were analyzed in parallel. Results The Atellica Hema 580 demonstrated acceptable imprecision within the manufacturer's specifications for whole blood and body fluid modes, good linearity for high and low ranges and no significant carryover. The full blood count, differential and reticulocyte correlated well with the Sysmex XN‐1000, except for mean cell hemoglobin concentration, basophil and large immature cells. The optical platelet count, reflexed in 34 samples with a platelet count <150 × 109/l, showed a strong correlation with the fluorescent platelet count on the Sysmex XN‐1000. The morphology flagging efficiency was 92% for white blood cells, 95% for red blood cells and 87% for platelets. Conclusion The Atellica Hema 580 showed good analytical performance and workflow efficiency for a wide range of patient samples.
hematology
What problem does this paper attempt to address?