Evaluation of haematology analyzer CELL-DYN 3700 SL

Enver Suljević,Muhamed Fazlić,Jozo Čorić,Emina Kiseljaković
DOI: https://doi.org/10.17305/bjbms.2003.3553
2003-05-20
Bosnian Journal of Basic Medical Sciences
Abstract:Research on the parameters of full blood count and differential white blood count is included in the program of all medical laboratories of primary, secondary and tertiary health care levels. Today, all haematological tests are exclusively performed on the haematology analyzers. Automation of haematology laboratories is a result of the huge requires for haematological test performing, timely issuing of the haematological findings, and possibility of the usage of modern techniques.This work is an evaluation of laser haematology analyzer Cell-Dyn 3700 SL. It investigates the reliability of test results throughout the following parameters: precision, accuracy, sensitivity and specificity of determination methods. It also explores the influence of sample transferring and correlation with haematology analyzer MAXM Retti. Haematology parameters that have been investigated are: white blood cell (WBC), neutrophils (NEU), lymphocytes (LXM), monocytes (MONO), eosinophils (EOS), basophils (BASO), red blood cells (RBC), haemoglobin (HGB), haematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular haemoglobin (MCHC) red cell distribution width (RDW), platelet (PLT), mean platelet volume (MPV), plateletocrit (PCT), and platelet distribution width (PDW).The results confirms that precision of analyzer fulfils the reproducibility of testing parameters: WBC, RBC, HGB, MCV, MCH, MCHC, and PLT. Correlation coefficient values (r) gained throughout the statistical analysis, that is linear regression results obtained throughout the comparison of two analyzers are adequate except for MCHC (r = 0.64), what is in accordance with literature data.Accuracy is tested by haematology analyzer method and microscopic differentiating method. Correlation coefficient results for granulocytes, lymphocytes and monocytes point the accuracy of methods. Sensitivity and specificity parameters fulfil the analytical criteria.It is confirmed that haematology analyzer Cell-Dyn 3700 SL is reliable for the determination of full blood count in everyday work. Analyzer and its program for differential white blood count can be used for the research and separation of normal and pathological blood counts with addition of microscopic methods confirming distribution or morphologic changes of leukocytes.
medicine, research & experimental
What problem does this paper attempt to address?