Performance evaluation of Bayer ADVIA 2120 automated hematology analyzer

WU Xin-zhong,PANG Xin,LUO Fu-dong,CHEN Fu,XIONG Xiao-quan
2007-01-01
Abstract:Object:To evaluate the performance of Bayer ADVIA 2120 automated hematology analyzer initially.Methods:The main parameters of instrument was evaluated by the guideline of the International Council for Standardization in Hematology(ICSH).Results:For the five main parameters of white blood cell(WBC),red blood cell(RBC),hemoglobin(Hb),mean corpuscular volume(MCV) and platelet(PLT),all CVs of total reproduction,within-run and between-day reproduction are below 4%,3% and 4%,carry-over are below 1.0%,and all have a more wide linearity range.Compare to the Coulter MAXM automated hematology analyzer,the correlation coefficient of them were 0.981,0.993,0.994,0.992 and 0.941 respectively.The screening capacity of immature granulocyte and abnormal lymphocyte is better than nucleated erythrocyte.Conclusion:Bayer ADVIA 2120 automated hematology analyzer has good feafures of precision,linearity,low carry-over,and strong screening capacity of abnormal cells.It's a perfect performance analyzer.