Study on setting up the reference method for the packed cell volume and its application

Li-ming PENG,Jia YANG,Li-xia BAI
DOI: https://doi.org/10.3969/j.issn.1007-4287.2007.05.021
2007-01-01
Abstract:Objective To set up the reference method for the packed cell volume(PCV).Methods The microhematocrit method for PCV was established according to the NCCLS' guideline and the performance of the method was evaluated;meanwhile the results of hematocrit(Hct) detected by Sysmex SE 9500 and 1800i were compared with the results of PCV obtained by microhematocrit method.Results The minimum centrifugation time was 4.5 min.There was no difference among the dimensions of capillary tubes(P>0.05).The within-run,between-run and overall coefficient of variance(CV) was less than 1%.There was minimal interference for PCV at high concentration of Hb(3g/L),triglyceride(TG=28.9mmol/L),bilirubin(DB=272.7mmol/L),and PLT(846×109/L)(P>0.05).Correlation between the SE 9500/1800i and reference method was excellent(r>0.975,P<0.001).Conclusion These finding demonstrate that the microhematocrit method for PCV is simple,precise,accurate,strong for anti-interference and it could be used to calibrate the Hct for CBC analyzer.