Accuracy evaluation of PLT-F method on low concentration of platelet counting

Juan CHEN,Jing DONG,Saihui PENG,Zhicheng WU
DOI: https://doi.org/10.3969/j.issn.1674-1129.2016.03.010
2016-01-01
Abstract:Objective To evaluate the accuracy of PLT-F channel on low platelet counting in the SysmexXN-9000 automated hematology analyzer. Methods Blood samples with low platelet concentration or those mixed with erythrocyte fragments were de-tected with SysmexXN-9000 automated hematology analyzer using three platelet count methods:PLT-F,PLT-I,and PLT-O. As comparison,the samples were labeled with CD61 antibody,and analyzed using flow cytometry. Results PLT-F method showed higher reproducibility and accuracy in comparison with PLT-I method and PLT-O method. Platelet count by PLT-F method were less affected by erythrocyte fragments than that by PLT-O method. There was a better correlation between the PLT-F system and the immunological reference method,despite of samples with low platelet concentration. Conclusions This study revealed that PLT-I method can accurately count platelets in the clinical routine blood samples. However ,blood samples with low platelet con-centration or those containing erythrocyte fragment should be reviewed by PLT-O or PLT-F methods ,even by microscopy or flow cytometry if necessary.