Use biological variation to value the lower detect limit of hematology and platelet aggregation test

WU Jun,ZHAO Jie,SONG Ying
DOI: https://doi.org/10.3969/j.issn.1007-4287.2013.05.015
2013-01-01
Abstract:Objective To detect the lower limit of hematology,platelet aggregation.Methods we use sysmex 2100 hematology analyzer blood cell account;Chrono-log platelet function instrument test platelet aggregation rate.After diluting samples,we retest the sample 10 times,plot X-axis with concentration,Y-axis with CV%,calculated the equation.We set 1/2 biological variation as standard CV to get the lower limit of detection.As for platelet aggregation,the concentration below the plateau is the lower limit concentration.Results Low limit detection of white blood cell,red blood cell,hematocrit,platelet is 0.11×109/l,0.14×1012/l,3.03%,15.31×109/l,respectivly.The low concentration of platelet aggregation is 200×109/l.Conclusion Use biological variation to set the lower limit of detection is practicable.It fit with clinical situation.The 1/2 biological variation could be the standard of lower limit of detection evaluation.