Comparison of peripheral blood automated hematopoietic progenitor cell count and flow cytometric CD34+ cell count

Mischa Hagenkötter,Thomas Mika,Swetlana Ladigan‐Badura,Karin Schork,Martin Eisenacher,Roland Schroers,Alexander Baraniskin
DOI: https://doi.org/10.1002/jca.22114
2024-05-07
Journal of Clinical Apheresis
Abstract:Background Stem cell apheresis in the context of autologous stem cell transplantation requires an accurate cluster of differentiantion 34 (CD34+) count determined by flow cytometry as the current gold standard. Since flow cytometry is a personnel and time‐intensive diagnostic tool, automated stem cell enumeration may provide a promising alternative. Hence, this study aimed to compare automated hematopoietic progenitor enumeration carried out on a Sysmex XN‐20 module compared with conventional flow cytometric measurements. Methods One hundred forty‐three blood samples from 41 patients were included in this study. Correlation between the two methods was calculated over all samples, depending on leukocyte count and diagnosis. Results Overall, we found a high degree of correlation (r = 0.884). Furthermore, correlation was not impaired by elevated leukocyte counts (>10 000/μL, r = 0.860 vs 20 000/μL, r = 0.843 vs <20 000/μL, r = 0.875). However, correlation was significantly impaired in patients with multiple myeloma (multiple myeloma r = 0.840 vs nonmyeloma r = 0.934). Summary Stem cell measurement carried out on the Sysmex XN‐20 module provides a significant correlation with flow cytometry and might be implemented in clinical practice. In clinical decision‐making, there was discrepancy of under 15% of cases. In multiple myeloma patients, XN‐20 should be used with caution.
hematology
What problem does this paper attempt to address?