The Comparison Research of Helical or Non-helical Scan at Convention Head CT

ZHOU Bing-chun,LI Ping,CAI Hong-fa,SHEN Xiao-jian
DOI: https://doi.org/10.3969/j.issn.1009-881X.2007.03.066
2007-01-01
Abstract:Objective To discuss the influnce of CT image quality between the way of helical and non-helical scan on the convention head CT inspection.Methods The head convention CT inspection through helical scan and non-helical scan was carried out on 30 volunteers by using Mx8000 4-row spiral CT with image grade on the resolution,the noise,the false shade and the synthesis,then a statistic comparison analysis was made.Results The CT images of helical scan on the false shade,the noise and the synthetic,are lower than that of the non-helical scan.There is an obvious difference between them(P<0.05);and there is little difference on the space resolution between them.Conclusion By using the way of non-helical scan,a comparatively highgrade CT image can be obtained.