Experimental study of low-dose cervical CT scan and its clinical application

陈海波,施丹,余日胜
2013-01-01
Abstract: Objective To access the optimal scan conditions of low- dose multi- slice CT on cervical region. Methods In the first step, 10 groups of low- dose scan conditions (140kV- 50mA, 140kV- 30mA, 120kV- 100mA, 120kV- 50mA, 120kV- 30mA, 90kV- 200mA, 90kV- 150mA, 90kV- 100mA, 90kV- 50mA , 90kV- 30mA) were selected according to the software prediction of CT radiation dose combined with the literature reports. In the second step, the necks of 3 goats were scanned with the conditions above. The quality of images was independently scored by two senior radiologists with double blind method and 9 groups of scan conditions (except 90kV- 30mA) met the requirement of diagnosis were selected. In the third step, 180 patients undergoing cervi-cal CT scan were randomly divided into 9 groups with 20 cases in each, patients were scanned with the conditions selected in animal experiments. And the quality of images was scored and anylyzed by SPSS 16.0 software. Results The CT radiation dose demonstrated a positive linear correlation with the change of tube voltage and current. The variety slope of CT radiation dose in-creased with the increasing of tube voltage or the current. The animal experiments demonstrated that 9 groups of conditions al met the requirement of diagnosis except 90kV- 30mA. The human cervical CT scan demonstrated that high radiation dose had a high score and bone scores were higher than soft tissue. 120kV- 100mA and 90kV- 200mA obtained good soft tissue images and excel ent bone images; 140kV- 50mA, 140kV- 30mA, 120kV- 50mA, 90kV- 150mA and 90kV- 100mA obtained good bone im-ages, and the diagnosis was not influenced by the soft tissue images. 120kV- 30mA and 90kV- 50mA obtained the soft tissue im-ages influencing the diagnosis and the bone images not influencing the diagnosis. Conclusion The images meeting the require-ment of diagnosis can be obtained by adopting low tube voltage and relative high tube current with lower radiation dose. In cervi-cal CT scan, 90kV- 100mA is the scan condition for soft tissue and 90KV- 50mA can be selected to observe the bone only.
What problem does this paper attempt to address?