Preliminary study on ultra low tube voltage (70kV)sequential scan with low-volume contrast media protocol for dual-source CT coronary angiography

曹剑,易妍,王怡宁,林路,王明,孔令燕,薛华丹,宋兰,王志伟,王少华,张龙江,卢光明,金征宇
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2014.06.003
2014-01-01
Abstract:Objective:To investigate the application of ultra low tube voltage (70kV)for coronary artery CT angiog-raphy (CCTA)with low-volume contrast media and low inj ection rate on dual-source CT (DSCT)equipment with integrat-ed circuit (IC)detector.Methods:Fifty patients with body mass index (BMI)not more than 25kg/m2 were scanned with a prospectively ECG-triggering sequential CCTA protocol on DSCT.All patients were randomly divided into two groups:the injection rate and volume of contrast in Group A was 4.0ml/s and 32mL (350mg I/mL);those in Group B were 4.0mL/s and 50mL (350mg I/mL),respectively.Tube voltage of both groups was 70kV.The raw data were reconstructed with sino-gram affirmed iterative reconstruction (SAFIRE)technique.The differences of age,BMI,heart rate,background noise,sig-nal-to-noise ratio (SNR)and radiation dose between the two Groups were compared.Besides,the CT value and CNR of the aortic root and each segment of coronary arteries were compared.And Image quality (1~4,excellent to non-assessable)be-tween the two Groups was compared segment-based.Results:There were no significant differences in age,BMI,heart rate, background noise,SNR or radiation dose between the two Groups (all P>0.05),and the mean ED was about 0.90mSv.The CT value of the distal segment of each coronary arteries in Group B was higher than Group A,while there were no signifi-cant differences in the proximal and median segments between the two groups.However,the CNR of each segment of coro-nary arteries in Group B were higher than Group A (all P<0.01).Mean segment-based image quality scores showed no sig-nificant difference between the two Groups reconstructed with SAFIRE (Group A,1.16±0.18,Group B,1.18±0.18;P=0.75).The percentage of assessable segments in each Group were more than 98%.Conclusion:For patients with normal BMI,the use of tube voltage as 70kV with SAFIRE technique in DSCT equipped IC detector is feasible.And both the con-trast injection rate and volume can be much lower.
What problem does this paper attempt to address?