Feasibility of low tube voltage protocol in lower extremity artery angiography with gemstone detectors CT

赵小英,赵金影,吴兴旺,陈坚,刘斌
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2015.01.019
2015-01-01
Abstract:Objective:To evaluate the feasibility of low tube voltage protocol for lower peripheral artery CTA using gemstone detectors CT.Methods:Totally 72 patients who were randomly divided in three groups(group A,B,C),underwent lower extremity CTA,the tube voltage as 120 kV,100kV,80 kV was used respectively,keeping the same tube current of 200 mAs.Other scanning parameters were kept identically.Contrast medium was injected automatically with bolu s tracking respectively.Image quality,segmental vascular enhancement,signal tonoise ratio(SNR),contrast to noise(CNR)and effective dose(ED)were independently evaluated in these 3groups.Results1 There was no significant difference of image quality of these 3groups.2The CT values of artey in group C was the highest and that in group A was the lowest There were statistical differences among three groups(P0.05).Statistics showed there was no significant differences in SNR and CNR among the three groups(P0.05).3Compared with group A,the CT DIvol,DLP and ED were reduced by35.8%,34.9%,34.4%respectively in group B and reduced by 65.8%,64.4%,64.4%respectively in group C.Conclusion:The tube voltage can be lowered from 120 kV to 80 kV for lower extremity CTA,in order to reduce the radiation dose,meanwhile the image quality and diagnostic confidence could be maintained.