Clinical eraluation of dual-source CT enterography using dual-energy rirtual non-enhanced technique

邓丽萍,史晓喆,章士正,张峭巍
DOI: https://doi.org/10.13609/j.cnki.1000-0313.2014.12.022
2014-01-01
Abstract:Objectire:To investigate the clinical feasibility of dual-source CT enterography using dual-energy virtual non-enhanced CT (VNCT)technique.Methods:CT enterography were performed in 39 patients with dual source CT scan-ner,including conventional non-enhanced CT scan (CNCT),arterial and venous phase contrast-enhanced CT scan (CECT) with dual energy mode (100kVp/230mAs and 140kVp/178mAs).VNCT images of arterial and venous phase were obtained after being post-processed with Liver-VNC software.Using CNCT as a standard,the image quality of VNCT images were evaluated and scored with 1~4 grade scoring system (the higher score indicated better image quality:4 scores means per-fect,1 score means unsatisfactory for diagnosis).The radiation dose of CNCT,arterial and venous phase CECT was com-pared.Results:Of 39 patients,there were 25 cases with lesions and 14 cases of normal.Compared with CNCT images,all the lesions were also detected accurately by VNCT images,and there was no significant difference of size and extent of lesions between CNCT and VNCT.The scores of VNCT images were all higher than 2,there was no significant statistic difference between the score of two doctors,and the image quality of VNCT images met the requirement of diagnosis.The average dose length product (DLP)of conventional non-enhanced CT,arterial and venous phase CECT were (283.8750±57.1195), (289.3750±53.1543 )and (301.2917 ± 55.3346 )mGy · cm respectively,with no statistic difference (F = 0.615,P =0.544).Conclusion:Dual energy CT virtual non-enhanced images may replace true non-enhanced images in CT enterography and then can reduce radiation dose of examination.
What problem does this paper attempt to address?