Effect of automatic tube current modulation range on radiation dose and image quality of head and neck CT angiography

华海琴,康德强,徐晓娟,李超,刘冰,刘明,傅瑶,王志群
DOI: https://doi.org/10.3969/j.issn.1005-8001.2017.01.011
2017-01-01
Abstract:Objective To investigate the effects of automatic modulation range of mA on the radiation dose and image quality of head and neck CT angiography (CTA) at the same voltage and noise index.Methods 29 patients underwent head and neck CTA using 120 kV tube voltage were randomly divided into one group at 150-600 mA and another group at 150-400 mA tube current modulation.The CT densities and noise of aortic arch,common carotid artery bifurcation,and middle of the basilar artery were measured.The ratio of signal to noise was calculated.The blood vessel visualization was scored on volume rendering,maximum density projection,and surface reconstruction images.The radiation dose parameters including CTDIvol and DLP as well as the tube current of the aortic arch,4th cervical vertebra body,dentate process,and the horizontal plane of the pituitary fossa were recorded.Results There was no significant difference in the qualitative scores and ratios of signal to noise of the aortic arch,common carotid artery bifurcation,and basilar artery in both groups.The CTDIvol and DLP of the 150-600 mA group were significantly higher than that of the 150-400 mA group.In both groups,the tube current was maxirnal at the level of aortic arch.C4 whereas the tube current was significantly higher in the odontoid process and the pituitary fossa of the 150-600 mA group than that of the 150-400 mA group.There was no significant difference between the two groups in the horizontal tube current of the pituitary fossa.Conclusion Selecting the optimal range of the tube current and effectively using automatic tube current adjustment technology may help to improve image quality and reduce the radiation dose of head and neck CTA.
What problem does this paper attempt to address?