Clinical evaluation of automatic tube voltage selection in chest CT angiography

Tilo Niemann,Simon Henry,Jean-Baptiste Faivre,Kanna Yasunaga,Sofiane Bendaoud,Ariana Simeone,Jacques Remy,Alain Duhamel,Thomas Flohr,Martine Remy-Jardin
DOI: https://doi.org/10.1007/s00330-013-2887-x
IF: 7.034
2013-07-05
European Radiology
Abstract:ObjectiveTo evaluate the clinical impact of automatic tube voltage selection on chest CT angiography (CTA).MethodsNinety-three patients were prospectively evaluated with a CT protocol aimed at comparing two successive CTAs acquired under similar technical conditions except for the kV selection: (1) the initial CTA was systematically obtained at 120 kVp and 90 ref mAs; (2) the follow-up CTA was obtained with an automatic selection of the kilovoltage (Care KV; Siemens Healthcare) for optimised CTA.ResultsAt follow-up, 90 patients (97 %) underwent CTA with reduced tube voltage, 100 kV (n = 26; 28 %) and 80 kV (n = 64; 69 %), resulting in a significant dose-length-product reduction (follow-up: 87.27; initial: 141.88 mGy.cm; P < 0.0001; mean dose reduction: 38.5 %) and a significant increase in the CNR at follow-up (follow-up: 11.5 ± 3.5 HU; initial: 10.9 ± 3.7 HU; P = 0.03). The increase in objective image noise at follow-up (follow-up: 23.2 ± 6.7 HU vs. 17.8 ± 5.1 HU; P < 0.0001) did not alter the diagnostic value of images.ConclusionAutomatic tube voltage selection reduced the radiation dose delivered during chest CT angiograms by 38.5 % while improving the contrast-to-noise ratio of the examinations.Key Points• As low a dose as possible must be used for CT angiography.• Automatic tube voltage selection permits reduced patient exposure.• Lowering the kVp enables increased intravascular attenuation.• Automatic tube voltage selection does not compromise the overall image quality.
radiology, nuclear medicine & medical imaging
What problem does this paper attempt to address?