The clinical observation of the effects of myopotential interference on the combipolar atrial sensing system and unipolar atrial sensing system in DDD pacemaker

许原,郭继鸿,刘肆仁,李学斌,黄卫斌,楚英杰
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.2001.02.006
2001-01-01
Abstract:Objective The aim of this study is to evaluate the incidence rate of the oversensing and various effects of myopotential interference on different sensing mode [combipolar atrial sensing(CAS) and unipolar atrial sensing(USA)] in DDD pacemaker,so to improve the understanding of the combipolar atrial sensing system.Methods Myopotential interference tests were performed in 30 patients with CAS and 30 patients with UAS during right arm isotonic exercise,all leads were implanted via right subclavian vein.Results There were no oversensing events when combipolar atrial sensing sensitivity was 0.5 mV.The oversensing occurred in 81.7% and 21.7% when unipolar atrial sensitivity was programmed to 0.5 mV and 1.0mV respectively.Pacemaker mediated tachycardia occurred in 50% patients with unipolar atrial sensitivity of 0.5 mV.Conclusion Compared with unipolar atrial sensing system,the DDD pacemaker with combipolar atrial sensing system has stronger ability of anti myopotential interference and longer lifespan of the generator.