The Clinical Observation of Auto AV Search Algorithm in Dual Chamber Pacemaker.

CHEN Jian-ming,ZHENG Liang-rong,WANG Zhi-jun
DOI: https://doi.org/10.3969/j.issn.1009-816x.2008.03.003
2008-01-01
Abstract:Objective To observe the effect of reducing ventricular pacing and rapid atrial arrhythmia with the auto AV search algorithm.Methods 43 Patients implanted with Medtronic kappa 700/900 DDD(R) pacemaker were randomized to the auto AV search algorithm or DDD/R mode for 3 month and then crossed over to the alternate pacing modality for additional 3 months.On the completing date of the crossover phase,the percent of ventricular pacing,atrial high frequence event and plasma BNP level were recorded during the first period and the second period of follow-up visits.Results 43 patients successfully completed the crossover study.Compared with the DDD/R mode,the cumulative percent of ventricular pacing,the atrial high frequence event and plasma BNP level were all significantly reduced in the auto AV search algorithm(15.9±4.8% Vs 78.0±10.3%,29±13 vs 58±15 and 65±21 pg/ml vs 333±30pg/ml,respectively.P<0.01).Conclusions The Search AV algorithm in the pacemaker effectively promotes intrinsic ventricular activation and substantially reduces unnecessary ventricular pacing.
What problem does this paper attempt to address?