Clinical Application and Long-Range Stability Observation of Single Lead VDD Pacing System

臧红云,王冬梅,韩雅玲,孙鸣宇
DOI: https://doi.org/10.3969/j.issn.1007-2659.2005.02.004
2005-01-01
Abstract:A single lead VDD pacing system was implanted in 25 patients with normal sinoatrial node function by subclavian venous puncture. The rate of atrioventricular synchronic pacing was observed by holter monitor during follow-up.Results:VDD pacing system was successfully implanted in 25 patients. The mean amplitude of P wave was 2.43±0.75 mV at implantation, and was significant higher than that of follow-up,P0.01. The rate of atrioventricular sychronization pacing was 99.97%±0.01%,99.98%±0.01%,99.98%±0.02%,99.99%±0.01% at 1 week,3months,6months,12months respectively. Postoperative complications such as atrial hyposensing, hypersensing and pacemaker mediated tachycardia were not observed. Conclusion: A single lead VDD pacing system is an optimal treatment for patients with high degree AVB and normal sinoatrial node function, which can be easily implanted and maintain steadily atrioventricular sychronization pacing.