Long-term Safety and Feasibility of Permanent His Bundle Pacing

Sheng Xia,Pan Yiwen,Zhang Jiefang,Ye Yang,Jiang Dongmei,Yang Ying,Sun Yaxun,Zhang Zuwen,Chen Shiquan,Jiang Chenyang,Fu Guosheng
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.2018.02.003
2018-01-01
Abstract:Objective The purpose of this study was to explore the long-term safety and clinical outcomes of permanent His-bundle pacing(HBP). Methods We consecutively recruited 105 patients scheduled to have permanent HBP implantation from January 2013 to December 2017 at Sir Run Run Shaw Hospital,Zhejiang University College of Medicine and explored the parameter changes and clinical outcomes. Results Eighty-eight patients received permanent HBP successfully(64 with selective HBP and 24 with non-selective HBP). In permanent HBP,the acute pacing threshold was(1.3±0.7) V/0.4 ms,and the sensed potentials were(4.1±3.5) mV,and the impedance was(582.7±157.6) Ω. During median follow-up of 15 months,the threshold and the sensed potentials had no significant difference compared with the measurements at implantation respectively. Median NYHA functional class was improved compared with pre-implantation. left ventricular ejection fraction(LVEF)increased from baseline in patients with reduced LVEF(n=33)while no significant change was found in patients with preserved LVEF. The rehospitalization rate was 15.9%(n=14), with 11 patitents due to heart failure. Conclusion Permanent HBP is feasible and safe for patients with a standard pacemaker indication,particularly for the patients with systolic heart failure.