T wave sensing test during ICD implantation procedure in inherited arrhythmias patients

Yaxun Sun,Jinhua Jin,Qin Lu,Qiang Liu,Zuwen Zhang,Chenyang Jiang,Guosheng Fu
DOI: https://doi.org/10.3877/cma.j.issn.2095-6568.2014.1.009
2014-01-01
Abstract:Objective To evaluate the T wave sensing in inherited arrhythmias patients during ICD implantation procedure .Methods In ICD implantation or replacement procedure , the T wave amplitude was measured by i-vitro ICD connection .The sensing threshold was programmed to the most sensitive to re-veal the possible TWOS .The threshold of ventricular sensing was plotted according to algorism of different models of ICD .Results Five Brugada syndrome patients , 1 long QT syndrome patient and 1 short QT syn-drome patient were enrolled in the study .With sensing threshold of 0.15mV in Virtuoso and Consulta ICD , TWOS were observed in 2 Brugada syndrome patients and the short QT syndrome patient .With sensing threshold of 0.3mV, TWOS was observed in Consulta ICD intermittently .Epic ICD showed TWOS under most sensitive mode in the short QT syndrome patient .Fortify, Vitality and Teligen showed no TWOS under most sensitive mode in all patients .No TWOS or inappropriate therapy was observed in 8-month follow-up. Conclusion T wave sensing could be evaluated in ICD implantation procedure and might help in reducing TWOS and inappropriate therapy.