Incidents and Predictors of Pre-Implantation QRS-T Morphology Screening in Subcutaneous Implantable Cardioverter Defibrillator

Wang Qingqing,Sun Zhe,Cao Feixia,Hu Lan,Su Yangang,Ge Junbo
DOI: https://doi.org/10.3760/cma.j.issn.1007-6638.2017.03.009
2017-01-01
Abstract:Objective To study the incidents of pre-implantation QRS-T morphology screening(TMS) in subcutaneous implantable cardioverter defibrillator(SICD)-eligible Chinese patients and predictors for TMS ineligible patients.Methods TMS was conducted in 80 consecutive outpatients and inpatients [83.8% male,age(57±13) years] who already had tansvenous implantable cardioverter defibrillators(TV-ICD) or needed ICD for primary or secondary preventions in Zhongshan Hospital,without an indication for cardiac pacing or accepting cardiac resynchronization therapy,using an ECG simulating the 3 sensing vectors of the SICD(TMS-ECG).Patients were defined suitable when at least 1 sensing vector was considered appropriate in supine,squat and standing position.Clinical,echocardiographic and standard ECG characteristics of patients who failed the test were compared to those who passed.Results In total,91.2% of patients were eligible for a SICD,8.8% of patients were considered not eligible according to the TMS-ECG.Seven patients failed TMS-ECG with the conventional left parastemal electrode configuration in all three subcutaneous ECG vectors,and one of them passed the TMS-ECG with the conventional right parasternal electrode configuration.Only the prolong QRS duration(≥ 150 ms,OR =0.017,95 % CI 0.001-0.416) was the independent predictors of failure in TMS.Conclusions We should be in special caution when selecting patients for SICD,especially the patients with wide QRSd,because their ECG may not be adequately sensed by SICD.
What problem does this paper attempt to address?