Abstract:Objective:To analyze the application of subcutaneous implantable cardioverter defibrillator (S-ICD) in China, evaluating the safety and efficiency of this device in preventing sudden cardiac death (SCD) .Methods:One hundred and eleven patients who underwent S-ICD from November 26, 2016 to December 11, 2019 were included and retrospectively analyzed. The baseline data, demographics, implanting characteristics (indications, site, pacing parameters and procedure times etc.) and follow-up detailed information (device interrogation etc.) were analyzed.Results:A total of 111 patients with full implanting and follow-up data from 42 centers in China were finally analyzed. Among them, 89 (89/111, 80.2%) were male. The mean age and the mean left ventricular ejection fraction were (47.2±16.6) years and 45.0%±16.6%, respectively. There were 39 cases (39/111, 35.1%) of primary prevention and 72 cases (72/111, 64.9%) of secondary prevention. The mean procedure time was (80.2±27.9) min. No severe complications were observed during the procedure. After median follow-up of 158 (54, 362) d, 75 times of defibrillation therapy were occurred in 27 cases. Of them, 22 (22/106, 20.8%) cases had 51 times of appropriate shock and 9 (9/106, 8.5%) cases had 24 times of inappropriate shock. The main reason of inappropriate shock was T-wave oversensing.Conclusion:The initial follow-up results of patients with S-ICD in China fulfilled anticipated goals for safety and efficacy.